Search Close Search
Search Close Search
Page Menu

Beckman Avanti J-26 XP

Beckman Avanti J-26 XP  Beckman Avanti J-26 XP      location: AS8-East wing rotor cabinet     status: in service
For tubes or microplates. Can take up to 24 plates per run. Features 2 acceleration and 3 deceleration rates, a delay start function and run times up to 99 hours and 59 minutes. Maximum Speed: 26,000 rpm Maximum RCF: 82,000 g.
Special feature: Uses JLA 8.1 rotor which holds 6-1 liter tubes. (see below)

Automatic Rotor Identification
• Dynamic Rotor Inertia Check — means no rotor can be run above its rated speed
• Temperature control is optimized based on the automatically identified rotor —no more too cold or too warm samples

Specifications
• Fully compatible with the J-Lite rotors series
• Fully adaptable with more than 32 rotor combinations (fixed angle and swinging bucket options ranging from 43.2 mL to 6 L capacity).
 
• Maximum Speed/g-Force26,000 rpm/82,000 x
• Accel/Decel Rates 2/3 
• Set Temp Range (1° increments)-10 to 40°C 
• Temp Control+ 2° of set temperature
• Speed Control + 10 rpm of set speed or 0.1%, whichever is greater
• Ambient Temp Range16° to 38°C

Rotors Availavle in MaPS dept. for this centrifuge
JS 5.3        With Microplate Kit, Swinging-Bucket, Aluminum, 24 x 96-Well Plates, 5,300 rpm, 6,130 x g
JLA 8.1     Fixed Angle, Aluminum, 6 x 1,000 mL, 8,000 rpm, 15,970 x g
JLA 16.25  Fixed Angle, Aluminum, Single Lock Lid, 6 x 250 mL, 16,000 rpm, 38,400 x g
JA 10        Fixed Angle, Aluminum, 6 x 500 mL, 10,000 rpm, 17,700 x g
JA 12        Fixed Angle, Aluminum, Single Lock Lid, 12 x 50 mL, 12,000 rpm, 23,200 x g

Documents to download:
Beckman Avanti J-26 userguide   Beckman Centrifuge Safety.pdf